Tabs – Explanation of Tabs on the Monitor
A. "Status"
View current alarms, which can be acknowledged or forwarded.
Deactivate an alarm for three pre-selected time intervals: 30 minutes, 4 hours, and 24 hours.
Use deactivation for situations where the cabinet is intentionally open for extended periods, such as restocking or cleaning.
B. "Overview"
Provides a complete overview of the measuring point, unit information, room information, alarm settings, regular contact persons, technical contact persons, and backstop function.
Changes can only be made by contacting Logit IQ.
C. "Graph"
Displays temperature measurements with a green graph indicating fluctuations.
Upper and lower alarm limits are shown as straight red lines.
The door status is indicated on the graph: a green dot when closed and a red dot when open.
Temperature at specific times is displayed.
Alarms are indicated with a triangle: a red triangle requires handling, a blue cross indicates the alarm has been acknowledged, and a green checkmark shows everything is confirmed OK.
D. "Data"
Shows the same information as the graph but in numerical and descriptive form.
Includes additional functionality in the upper right corner to display average, maximum, and minimum temperatures for the selected period.
Under the "Graph" and "Data" tabs, you have four options:
1. "Show Active Alarm" – only displayed if there is an active alarm.
2. "Show Last Hour"
3. "Show Last 24 Hours"
4. "Show Last Week"
E. "Alarm History"
Displays all alarms that have occurred on the unit.
Press the plus button for more details.